Patient Monitor User Manual Alarm Information
- 50 -
Message Cause Alarm Level
RR High
RR measuring value is above upper alarm limit. User-selectable
RR Low
RR measuring value is below lower alarm limit. User-selectable
SpO
2
High
SpO
2
measuring value is above upper alarm limit.
User-selectable
SpO
2
Low
SpO
2
measuring value is below lower alarm limit.
User-selectable
SpO
2
No Pulse
The signal of the measurement site is too weak due to
insufficient blood supply and environmental factors,
so the monitor can’t detect the pulse signal.
High
SpO
2
Desat
SpO
2
measuring value is below the SpO
2
Desat Limit.
High
PR High
PR measuring value is above upper alarm limit.
User-selectable
PR Low
PR measuring value is below lower alarm limit. User-selectable
T1 High
Measuring value of T1 channel is above upper alarm
limit.
User-selectable
T1 Low
Measuring value of T1 channel is below lower alarm
limit.
User-selectable
T2 High
Measuring value of T2 channel is above upper alarm
limit.
User-selectable
T2 Low
Measuring value of T2 channel is below lower alarm
limit.
User-selectable
TD High
Measuring value of TD channel is above upper alarm
limit.
User-selectable
SYS High
SYS measuring value is above upper alarm limit. User-selectable
SYS Low
SYS measuring value is below lower alarm limit. User-selectable
DIA High
DIA measuring value is above upper alarm limit. User-selectable
DIA Low
DIA measuring value is below lower alarm limit. User-selectable
MAP High
MAP measuring value is above upper alarm limit. User-selectable
MAP Low
MAP measuring value is below lower alarm limit. User-selectable
PR (NIBP) High
PR measuring value from the NIBP module is above
upper alarm limit.
User-selectable
PR (NIBP) Low
PR measuring value from the NIBP module is below
lower alarm limit.
User-selectable
Art SYS High
Art SYS measuring value is above upper alarm limit. User-selectable
Art SYS Low
Art SYS measuring value is below lower alarm limit. User-selectable
Art DIA High
Art DIA measuring value is above upper alarm limit. User-selectable
Art DIA Low
Art DIA measuring value is below lower alarm limit. User-selectable
Art MAP High
Art MAP measuring value is above upper alarm limit. User-selectable