Patient Monitor User Manual Alarm Information
- 56 -
Message Cause Alarm Level Action Taken
RESP Cardiac
Artifact
No RESP waveform can be
detected due to apnea or
shallow breathing of the
patient.
High
Check whether the
patient is breathing
normally. Take
measures to help the
patient breathe
normally when
necessary. If the
patient is breathing
normally, try to
adjust the electrode
position on the
patient in order to
reduce the
interference of
cardiogenic artifact.
SpO
2
SpO
2
Sensor Off
SpO
2
sensor may be
disconnected from the
patient or the monitor.
Low
Make sure the
sensor is well
connected to the
patient’s finger or
other parts. Make
sure the monitor
and cables are well
connected.
SpO
2
Comm Fail
SpO
2
module failure or
communication failure
High
Stop using
measuring function
of SpO
2
module,
and notify
biomedical engineer
or manufacturer’s
service staff.
SpO
2
No Sensor
No SpO
2
sensor was
connected to the monitor.
Low
Make sure the
monitor and sensor
are well connected,
reconnect the
sensor.
SpO
2
Sensor Err
Malfunction in the SpO
2
sensor or in the extension
cable.
Low
Replace the SpO
2
sensor or the
extension cable.