Patient Monitor User Manual Alarm Information
- 64 -
Message Cause Alarm Level Action Taken
CO
2
Occlude
Water trap of SideStream is
occluded.
High
Make sure the gas
exhaust works well
BIS
BIS Comm Fail
1) Disconnection between
the V-BIS module and BISx
device.
2) BISx device stops
operating.
High
Properly connect
cables and well
connect the module.
BIS Sensor Not
Connected
1) The sensor is not
properly connected.
2) PIC is not properly
connected.
Low
Reconnect the
sensor or PIC.
BIS Sensor Type
Error
1) Wrong sensor type.
2) Use the sensor on
neonatal patients.
Low Replace the sensor.
BIS Sensor Usage >
24hrs
The sensor was attached to
the monitor for more than
24
hours.
Low Replace the sensor.
BIS Sensor Error
Sensor malfunction
including sensor over
current, sensor ground
element (positive and
negative) failure.
Low
Examine sensor
connection or
replace the sensor.
And then click
Continue in the
BIS Sensor Fault
window which
appears on the
screen or reconnect
the V-BIS module.
BIS Sensor Invalid
1) The BIS sensor is invalid
or not supported by the
BISx device.
2) The sensor is not
properly connected
Low
1) Replace the
sensor.
2) Connect the
sensor properly.