5-30 Remote Operation
MMEMory:INITialize
Sending the command clears the entire contents
of
the internal RAM.
MMEMory:LOAD[:SETup]
<setup_name>
Sending this program message recalls the stored instrument setup setup_name
from memory. The setup_name must contain the file extension .SET to correctly
identify the file.
This is the equivalent to using the IEEE-488
*
RCL
command, except the
<setup_name>
is
a
character
string
rather
than
a
numeric
value.
MMEMory:LOAD[:SETup] <setup_name> can load a file created by *RCL by
internally supplying an "S" prefix. For example:
*RCL 172345
is equivalent
to:
MMEM:LOAD:SET S172345
MMEMory:STORe[:SETup]
<setup_name>
Sending this program message stores the
CUITent
instrument settings as
"setup_name.SET"
10
the internal memory. This the equivalent
10
using the IEEE-
488 *SAV command, except <setup_name> is a character string rather than a
numeric value. For example:
*SAV32
is equivalent
to:
MMEM:STOR:SET S32
MMEMory:DLOad[:DSO] <DSO_name>
Sending tbis program message uploads the DSO driver "DSO_name.DSO" in to
Mode1395 memory.
For
details
on
the message format, refer to appendix F.