TRIGger:SOURce[:STARt]
<INTernal
1
EXTernal
1 MANual>
Sending this program message selects the source
of
the start trigger signal. The
start trigger signal initiates activity when the instrument is in a triggered mode
of
operation.
EXTernal
Selects an external signal
jack
as the trigger source.
INTernal
Selects an instrument dependent internal signal as the trigger source.
MANual
Selects the MAN TRIO key as the trigger source.
TRIGger:TIMer
<numeric_value>
(le-3)
Sending this program message sets the period
of
an internal periodic signal source.
The timer signal acts as a trigger when
it
is the selected trigger source.
It
ranges
from 2e-7 to
le4
seconds with 2e-7 resolution.
Remote Operation 5-49