EasyManua.ls Logo

GE LOGIQ Totus

GE LOGIQ Totus
644 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
Section 8-8 - Replacing Main Console 8 - 147
8-8-19 Replacement of the T2 Handle
This table includes information specific to these instructions. For more information 8-2-6 "Tools needed
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
8-8-19-1 Preparations
1.) Remove the T2 Side Cover L Assy as described in 8-8-7 "Replacement of the T2 Side Cover L
Assy" on page 8-135
2.) Remove the T2 Side Cover R Assy as described in 8-8-8 "Replacement of the T2 Side Cover R
Assy" on page 8-136
3.) Remove the Rear Cover Mid Assy-T as described in 8-8-10 "Replacement of the Rear Cover Mid
Assy-T" on page 8-138
4.) Remove the Rear Cover Top Assy-U as described in 8-8-15 "Replacement of the Rear Cover Top
Assy-U" on page 8-143
8-8-19-2 Removal Procedure
8-8-19-3 Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8-
12 on page 8-223
Table 8-81 Manpower, Time, Tools, and PPE
Manpower Time Tools PPE
One person Approximately 10 minutes
Phillips Screwdriver
Hexagonal (Allen) Wrench
Cut resistant gloves
Knee Pad
Step Corresponding graphic
1.
Unscrew 4 screws to remove the Unity Cable Cover
Brkt from the console.
2.
Unscrew 6 screws to remove the T2 Handle from the
console

Table of Contents