EasyManua.ls Logo

GE Voluson P6 - 3-8-3 On-Board Optional Peripherals

GE Voluson P6
553 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
3-60 Section 3-8 - System Configuration
3-8-3 On-Board Optional Peripherals
Mains outlet: AC outlet for on-board peripherals
Output voltage : 100~250V.
CAUTION
!! CAUTION:
Modification of voltage setting only by an authorized service person!
The maximum power consumption of equipment (inclusive of the color LCD monitor) connected
to these outlets must not exceed 400VA!
Table 3-8 Approved Peripherals
Device Manufacturer Model Connection Comment
Digital B/W printer SONY UP-D897 USB-Port
Digital Color Printer SONY UP-D25MD USB-Port
Digital Deskjet Printer
(Bluetooth)
HP Officejet H70 USB-Port (Bluetooth)
DVR GEUK Dedicated Connector
ECG Module Norav Marlin/ECGUSB1D USB-Port
Footswitch Whanam Elec. USB-Port