GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-87
Section 8-40
Replacement of the Caps for Hardkeys
8-40-1 Replacement of Circle Key Caps only
8-40-1-1 Manpower
One person, approx. 1 minute/cap
8-40-1-2 Tools
Small-sized slotted screwdriver or tweezers.
8-40-1-3 Circle Key Caps - Replacement Procedure
1.) By means of a small slotted screwdriver, carefully push against the circle hardkey cap.
2.) Lift the cap, until it is completely loosened from its base.
3.) Place the new hardkey cap down until it snaps into position.
NOTICE
Please observe that replacement procedure depend on key caps that have to be replaced!
• if just the circle key caps have to be replaced.
• if trackball buttons or mode key slices have to be replaced too, please contact your local distributor
or GE service representative.
Figure 8-104 Push against the circle cap and lift it