GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
7-10 Section 7-4 - Screen Captures and Logs
7-4-1-2 How to use a Screen Filter
In addition to the “pre” capture filter, use the Filter tool in the screen to filter what is displayed.
Filter the DICOM packets, since they are the most probable for the troubleshooting.
1.) Type dicom in the “Filter” field and click on APPLY
.
The display should filter all DICOM packets, filtering out image data.
2.) To deactivate the filter, click on CLEAR
.
Figure 7-9 Filter: dicom
Filter
Apply