GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
8-54 Section 8-24 - Replacement of the SIDE TRAY
Section 8-24
Replacement of the SIDE TRAY
8-24-1 Manpower
One person, 2 minutes
8-24-2 Tools
Philips screwdriver
8-24-3 Preparations
1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
8-24-4 Removal Procedure
1.) Unscrew 2 screws and remove the SIDE TRAY. Refer to the figure below.
8-24-5 Installation Procedure
1.) Install the new parts in the reverse order of removal.
2.) Perform: FRU8-25: Replacement of the SIDE TRAY - Functional Tests.
Figure 8-65 Unscrewing 2 screws and remove SIDE TRAY ASSY