GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013)  PROPRIETARY SERVICE MANUAL
5-52 Section 5-13 - Service Page
Section 5-13
Service Page
5-13-1 Introduction
The Service Page contains specific software/hardware test modules, system setup, update, etc. for 
Voluson systems only.
5-13-2 Access / Security
The service page has different access and security user levels.
Each user is only granted access to the tools that are authorized for their use.
5-13-3 Service Login
1.) Press the F2 key on the control panel.
2.) In the “Utilities” menu select SYSTEM SETUP
 to invoke the setup desktop on the screen.
3.) If not already done, connect the USB “standard GE” service dongle to the USB connector(s) on the 
rear of the system. As soon as the dongle is detected the login window pops up.
4.) Enter the “Standard GE Revolving” password (changes every six months) and then click ENABLE.
5.) On the left side of the screen select ADMINISTRATION and then click the SERVICE tab.
6.) Click the SERVICE TOOLS
 button on the Service window to get access to the “Service Tools” page.
5-13-3-1 Auto Tester
Autotester is a log function of customer activities. It records all user actions (scanning, entries, 
performing Calculations, review of Patient Reports, etc.…). It is possible to safe (record) as file on HDD. 
But also export to DVD/CD can be done to allow replay of the records on other units.
Operation see: Section 7-7 "How to use the Auto Tester program" on page 7-27.
Figure 5-33   Service Tools window
NOTICE
!! NOTICE:
For intermittent problems this file can be requested from the Service Engineer or customer.
It is possible to burn this file on DVD/CD+R/RW.
Export System 
Auto Tester
Monitor Test
Delete all patients
Update
TCP/IP Buffersize
Common
Keyboard Layout
Service Desktop
Exit
Data
Request for
Service
Load Application
Settings