GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
7-56 Section 7-12 - Troubleshooting Trees, Instructions and Tech Tips
7-12-5 Monitor Troubleshooting
For further details refer to Section 6-3 "LCD Monitor Adjustment" on page 6-2.
Fault symptom Check these items
No image/No signal
Check the main panel cable is properly connected to Monitor.
Check the main panel cable is properly connected to RFS.
Check no pins of the main panel cable are bent.
Video test pattern are not clear, bright
parallel or square
Replace the monitor.