Chapter 5 - Components and Functions (Theory) 5-61
5-15-5 Media Menu
1.) If system functionality allows it, collect vital system information as described in 5-15-3 "Before
Recovering or Initializing the Hard disk" on page 5-58.
2.) Insert the System DVD into the DVD/CD+R/RW Drive.
3.) In the “Voluson Maintenance System” main menu (Figure 5-38) select [Media].
The “Media Menu” with following functions appear:
5-15-5-1 Exit
Return to the “Voluson Maintenance System” main menu.
5-15-5-2 Restore
This function takes the image stored on the “System DVD” to replace the existing image of the “C”
partition. Additionally the (R:) Rescue image of (C:) is replaced by the “DVD image”.
The System “C” image contains:
• Windows XP with patches
• Ultrasound Application Software [UIS]
• Image (Application) Settings [Backup]
• Electronic User Manual [EUM]
NOTICE
The “Media Menu” is only accessible if the System DVD (see: Table 9-13) is inserted into the system’s
DVD/CD+R/RW Drive
Figure 5-39 Media Menu
NOTICE
Data on the D: [User] partition (e.g., Patient Archive images, Service Database, Measurements,
Patient Data and Reports) will NOT be modified!
NOTICE
If the currently installed software has to be upgraded to a newer version, and the system DVD being
used is SW 13.0.0 or higher, a new software specific “Permanent key” is required.
Restore
Initialize
New Disk
Only available when
booting from DVD