GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-45
7-9-6-7-2 Tx Signal Path Test
It verifies Tx Signal from TxPulser on BF board by firing test Tx pulse wave and reading back from
corresponding Rx channel, and displays each channels diagnostic result, and the overall test result as
PASSED/FAILED.
7-9-6-7-3 CW Signal Path Test
It verifies CW Signal Path from AFE to RFS by generating test wave signal and recording it with I Q
phase recorder, and displays each channels diagnostic result, and the overall test result as PASSED/
FAILED.
7-9-6-8 ECG Overall Test
This diagnostic test performs the self-diagnosis of ECG interface.
Table 7-9 below outlines Voluson® P8 / Voluson® P6 Diagnostics Tool described in the sub-sections.
7-9-6-8-1 ECG Interface Test
It verifies connection status and trace status on ECG board, and displays the connection status, trace
status, board version information and the test result as PASSED/FAILED.
Table 7-9 Automatic diagnostic test under ECG Test
Sub-section Description Page Number
7-9-6-8-1
ECG Interface Test
7-45