GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5-55
Section 5-14
Boot Screen Functions
5-14-1 Overview
Following LINUX supported functions are available as soon as the “Boot Screen” appears:
5-14-1-1 Voluson
The System will boot-up in windows. The Ultrasound Application is started.
For details refer to Section 3-6-2-3 "During a normal boot, you may observe" on page 3-26.
5-14-1-2 GE-Service
The Voluson Maintenance System functions are secured by service password (class C).
1.) Turn system OFF and then back ON.
2.) As soon as the “Boot Screen” appears (see: Figure 5-36 above), press the PG DN [ Arrow
down] key on the keyboard until the GE Service item is highlighted, then press ENTER.
3.) Enter the Password Rudi (case sensitive) and confirm with ENTER. The system boots into
LINUX.
Figure 5-36 Boot screen
NOTICE
!! NOTICE:
After 3 sec. without pressing any key, the system will boot-up in windows (= Voluson item).
CAUTION
After 3 sec. without pressing any key, the system will boot-up in windows (= Voluson item).
If you missed step 2, retry again with CTRL + ALT + DEL.
Voluson
GE-Service
Rollback
Memtest