GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-77
Section 8-35
Replacement of the CBP
8-35-1 Manpower
One person, 30 minutes
8-35-2 Tools
Philips screwdriver, Allen Wrench
8-35-3 Preparations
1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
2.) Remove the rear cover as described in Section 8-14 on page 8-33.
8-35-4 Removal Procedure
1.) Remove the CSI. Refer to the replacement of CSI.
2.) Unscrew 8 screws and remove fan connector.
Figure 8-92 Unscrewing 8 screws