8-62 Section 8-28 - Replacement of the RFS
8-28-5 Installation Procedure
1.) Install the new parts in the reverse order of removal.
2.) Perfom 8-28-6 Serial Number - Reprogamming Procedure to reprogram the serial number.
8-28-6 Serial Number - Reprogamming Procedure
Before calling OLC please prepare the following data:
SNO of Voluson® P8 / Voluson® P6 (Alphanumeric serial number can be found on main label at
back of system)
Hardware ID number (See SYSTEM INFO page or label of RFS-board)
1.) Gather System Serial number and Hardware ID
- Press the F2
key on upper OPIO, select SYSTEM SETUP, select ADMINISTRATION, click
on SYSTEM INFO
and write down the "Hardware ID" number.
NOTICE
!! NOTICE:
To reprogramming the System Serial Number, contact the OLC(On Line Center) to get the necessary
"Serial Key" and "Permanent Key" (calculated in OKOS) !!!
Figure 8-74 System Setup - Administrator - System Info page
NOTICE
!! NOTICE:
The Hardware ID can also be found on the label on the RFS-board.
Serial number
Hardware ID