GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013)  PROPRIETARY SERVICE MANUAL
4-44 Section 4-5 - Backup and Restore Database, Preset Configurations and Images
4-5-6 Archiving Images 
4-5-6-1 Save Image Archive
NOTE: A backup of the Image Archive always contains the Patient Archive (database containing patient 
demographic data and measurements) + images of the selected exams.
1.) Press the F2
 key on the control panel.
2.) In the “Utilities” menu select SYSTEM SETUP
 to invoke the setup desktop on the screen.
3.) On the left side of the screen select BACKUP
 and then click the IMAGE ARCHIVE tab.
4.) Click the SAVE
 button.
CAUTION
!! CAUTION: 
It is recommended to Save Image Archive (Section 4-5-6-1 on page 4-44) and Backup the Full 
System Configuration (Section 4-5-3 on page 4-39) once a week.
Figure 4-32  System Setup - Backup - IMAGE ARCHIVE page
Figure 4-33  Image Archive Save - preparations
[Save] button
[Load] button
click [Next] button
[Advanced] button
choose archiving
span time
check mark if desired