GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013)  PROPRIETARY SERVICE MANUAL
8-104 Section 8-42 - Replacement of FAN with Cable ASSY II
Section 8-42
Replacement of FAN with Cable ASSY II
8-42-1 Manpower
One person, 30 minutes.
8-42-2 Tools
Philips screwdriver, Allen Wrench
8-42-3 Preparations
1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
2.) Remove the rear cover as described in Section 8-14 on page 8-33.
8-42-4 Removal Procedure
1.) Unscrew 4 screws and remove fan connector.
8-42-5 Installation Procedure
1.) Install the new parts in the reverse order of removal.
NOTE: In case of installing FAN with Cable ASSY (GPN:5363484), replace screws to SCREW PH M4X30 
(GPN:5176333) in FRU.
2.) Perform:  FRU8-44: Replacement of FAN with Cable ASSY II - Functional Tests
Figure 8-133