EasyManua.ls Logo

GE Voluson P6 - Section 8-20 Replacement of the TGC BOARD; 8-20-1 Manpower; 8-20-2 Tools; 8-20-3 Preparations

GE Voluson P6
553 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-47
Section 8-20
Replacement of the TGC BOARD
8-20-1 Manpower
One person, 10 minutes
8-20-2 Tools
Philips Screwdriver
8-20-3 Preparations
1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
8-20-4 Removal Procedure
1.) Follow the removal procedure of Section 8-19-4 (step 1~3)
2.) Remove 8 TGC KNOBS. Refer to the figure below.
Figure 8-57 Removing 8 TGC KNOBS

Table of Contents