GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-73
Section 8-33
Replacement of the PIR
8-33-1 Manpower
One person, 30 minutes
8-33-2 Tools
Philips screwdriver, Allen Wrench, Hirose ejector
8-33-3 Preparations
1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
2.) Remove the rear cover as described in Section 8-25 on page 8-55, except for pulling out the NEST.
3.) Remove 3 RS Knobs.
8-33-4 Removal Procedure
1.) Draw the BF/PIR out of the Nest Frame.
Figure 8-86 BF/PIR
Figure 8-87 Drawing the BF/PIR
BF/PIR