4-38 Section 4-5 - Backup and Restore Database, Preset Configurations and Images
4-5-2-3 Load only parts of the “Complete Backup”
NOTE: Following procedure should be used, if the user has saved own User Programs, 3D/4D Programs or 
AutoText entries, but new presets have to be added to match the installed Application Software version 
(e.g., User Programs and 3D/4D Programs for new probe, etc.). 
1.) Perform Preparations on page 4-36. 
2.) Click the +
 sign (next to “Complete Backup”) to open the content tree. 
3.) Click the +
 sign next to “UserPrograms”, select the desired probe(s) and copy the content by clicking 
the >>
 button.
4.) Click the +
 sign next to “3D/4D Programs”, select the desired probe(s) and copy the content by 
clicking the >>
 button.
NOTE: To return selected items from the “Load Data” field to “Backup Data” field select the <<
 button.
5.) Confirm selection with the LOAD
 button.
Settings will be loaded and the US Application Software restarts.
Figure 4-23  Load image settings for probe (e.g., RAB4-8-D)
copy desired 
content into 
right field
[+] sign
>>
 button
[Load] button