Chapter 8 - Replacement Procedures 8-37
Section 8-15
Replacement of the AIR FILTER
8-15-1 Manpower
One person, 1 minutes
8-15-2 Tools
None
8-15-3 Preparations
1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
8-15-4 Removal Procedure
1.) Pull out the FRONT COVER ASSY. Refer to the figure below.
2.) Pull out the AIR FILTER. Refer to the figure below.
Figure 8-44 Removing the FRONT COVER ASSY
Figure 8-45 Removing the AIR FILTER