GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
8-122 Section 8-45 - Replacement of ECG Kit assy
FRU8-41 Replacement of FULL AN KEY
FRU8-42: Replacement of the Caps for Hardkeys
Functional and/or Leakage Current Test - N/A
FRU8-43: Replacement of FULL AN KEY
FRU8-44: Replacement of FAN with Cable ASSY II
FRU8-45: Functional Checks after FRU replacement (Debrief Guidelines)
See Section
Functional and/or
Leakage Current Test
Debrief Script
4-3-1
Power On / Boot Up Proprietary Service Manual,
Voluson® P8 / Voluson®
P6
,
Revision Voluson® P8 /
Voluson® P6, Section 8-39
Equipment passed all required
tests and is ready for use.
7-11-1
Console Test
Table 8-1
System Function
4-3-2
Power Off / Shutdown
See Section
Functional and/or
Leakage Current Test
Debrief Script
10-8-5
Enclosure Leakage Current Test Proprietary Service Manual,
Voluson® P8 / Voluson®
P6
,
Revision Voluson® P8 /
Voluson® P6, Section 8-41
Equipment passed all required
tests and is ready for use.
4-7
Peripheral Checks
See Section
Functional and/or
Leakage Current Test
Debrief Script
4-3-1
Power On / Boot Up Proprietary Service Manual,
Voluson® P8 / Voluson®
P6
,
Revision Voluson® P8 /
Voluson® P6, Section 8-42
Equipment passed all required
tests and is ready for use.
Table 8-1
System Function
4-3-2
Power Off / Shutdown