EasyManua.ls Logo

GE Voluson P6 - Page 269

GE Voluson P6
553 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
5-62 Section 5-15 - LINUX - Voluson Maintenance System
Operation:
1.) Select [Restore] in the displayed “Media Menu” (see: Figure 5-39 above).
NOTE: If the message: “Recovery system (Linux) needs to be updated ...” appears, confirm with YES
.
NOTE: If the currently installed software has to be upgraded to a newer version, and the system DVD being
used is SW 13.0.0 or higher, a new software specific “Permanent key” is required.
2.) Only if you are sure you have saved the “Full System Configuration”, select YES
in the
following dialog box (see: Figure 5-40 below) and confirm with ENTER
.
NOTE: If the message: “Found obsolete image files on R: .... “appears, select YES
and confirm with ENTER.
3.) Recovering will take some time.
Please wait until the final window (Figure 5-41 below) appears and then press ENTER
.
Figure 5-40 Confirm dialog with YES
Figure 5-41 Restore C: done