GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013)  PROPRIETARY SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5-59
- Click the GENERAL tab and print out or write down the settings of this page.
- Click the USER SETTINGS and print out or write down the settings of this page.
- Click the PATIENT INFO DISPLAY tab and print out or write down the settings of this page.
8.) Press the F2
 key on the control panel.
9.) In the Utilities menu, select MEASURE SETUP to invoke the measure desktop on the screen.
10.)Print out the MEASURE & CALC page or write down the settings of this dialog page.
Ask the customer if he is using special programmed tables or equations.
11.)Print out the APPLICATION PARAMETERS page or write down the settings of this dialog page.
12.)Print out the GLOBAL PARAMETERS page or write down the settings of this dialog page.
13.)Disconnect all external USB devices (e.g. USB printer).
NOTE: Do NOT disconnect the Disk Drive Module.