GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-47
4.) Click on CONTINUE. The "Probe Selection" menu appears on the screen.
5.) Press the FREEZE
key. The application starts, but does not show any Ultrasound information.
6.) Backup all images (see:Section 4-5-6 "Archiving Images" on page 4-44).
7.) Save a "Full Backup" as described in Section 4-5-3 "Save Full System Configuration (Full Backup)"
on page 4-39) or press the ALT+D
key to gather detailed information and data from the system.
(see: Section 7-4-2-1 "Export System Data (by pressing the ALT + D key)" on page 7-11).
8.) Press the F2
key on the control panel and select SYSTEM SETUP to invoke setup desktop.
9.) Select ADMINISTRATION
, open the SYSTEM INGO page and click REBOOT MACHINE.
Figure 7-49 H Key pressed during startup