GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
7-68 Section 7-13 - Error Messages
No CD Writer found
check the connection and the Power cable - plug the cable
off and on and try again. (restart the system)
No default settings available - use "Edit Settings" to set the Default Settings! select a printer
No disc or device not ready!
insert disk, if fails again reboot and try again (with another
disk)
No DVD in drive! Put DVD into drive of Recorder. Put DVD into drive
no PCI Interface restart the system
No Printer selected! select a printer
Not enough space. (Not enough space on destination to hold the backup data.) select another destination to save Full Backup
overview window creation failed. Action: restart the system restart the system
Password contains invalid characters. Allowed characters are A..Z, a..z, 0..9 and . Enter password without invalid characters
Password is too short. Minimum 6 characters. Enter password with at least 6 characters
Password must contain at least 2 non-letter characters, 0..9 or Enter password with minimum 2 non-letter characters
Persistance Out Of Range press ok and save this user-setting once again
Persistence coeff page index too big restart the system
pGetActualUnitBuffer failed, RepresentationManager is not initialized restart the system
Please plug off and on probe and try again
plug of and on the probe and try again, plug it on a different
probe connector.
Please plug off and on probe and try again. Plug it on a different probe connector see error message, connect another probe
pNextUnitCompleted failed, RepresentationManager is not initialized restart the system
PowerManager: HRS not initialized restart the system
PowerSpecBuffer memory allocation failed. Action: restart the system restart the system
PRF_GeneratorBoundary: BBC Ensemble Limitation out of limit.
Action: restart the system
restart the system
PRF_GeneratorBoundary: BBCPW Ensemble Limitation out of limit
Action: restart the system
restart the system
Probe () doesn't contain correct ProbeVersion! Action: Please contact online center Please contact technical support
Probe Scan Function Not Supplied restart the system
Probe with Probe-ID xx not supported. On Attempt to read Probe-ID from an invalid
probe connector. Action: restart system
Connect another probe
Probe with Probe-ID xx not supported on left probe connector (A). Connect another probe
Probe with Probe-ID xx not supported on middle probe connector (B). Connect another probe
Probe with Probe-ID xx not supported on right probe connector (C). Connect another probe
Probe-ID valid, but Flexible-EEPROM-Datamodel not implemeted yet.
On Attempt to read Probe-ID from an invalid probe connector. Action: restart system
restart the system