8-112 Section 8-45 - Replacement of ECG Kit assy
3.) Connect the USB cable with Keyboard assembly as shown.
4.) Assemble Keyboard assembly.
Figure 8-144 USB cable
5.) Fasten 9 numbers of fasteners below the Keyboard base.
6.) Assemble 4 probe holders.
Figure 8-145 Keyboard base and probe holders