GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
8-114 Section 8-45 - Replacement of ECG Kit assy
FRU8-5: Full Backup (Full System Configuration) - Loading Procedure
go to 4-5-4 Load Full System Configuration (Full Backup)
FRU8-6: Image Archive - Loading Procedure
go to 4-5-6-2 Load Image Archive
FRU8-7: Replacement or Activation of Options
FRU8-8: Replacement of Monitor
FRU8-9: Replacement of OPIO
See Section
Functional and/or
Leakage Current Test
Debrief Script
Table 8-1
System Function Proprietary Service Manual,
Voluson® P8 / Voluson®
P6
,
Revision Voluson® P8 /
Voluson® P6, Section 8-8.
Equipment passed all required
tests and is ready for use
functional checks depend on activated option
See Section
Functional and/or
Leakage Current Test
Debrief Script
4-3-1
Power On / Boot Up Proprietary Service Manual,
Voluson® P8 / Voluson®
P6
,
Revision Voluson® P8 /
Voluson® P6, Section 8-9.
Equipment passed all required
tests and is ready for use.
6-3
LCD Monitor Adjustment
Table 8-1
System Function
4-3-2
Power Off / Shutdown
See Section
Functional and/or
Leakage Current Test
Debrief Script
Table 8-1
System Function
Proprietary Service Manual,
Voluson® P8 / Voluson®
P6
,
Revision Voluson® P8 /
Voluson® P6, Section 8-11
Equipment passed all required
tests and is ready for use.