GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
Chapter 10 - Care & Maintenance 10-17
Table 10-18 Type CF Patient Auxiliary Limits
Table 10-19 Isolation between Mains and Applied Limits****- Patient Leakage Main to Type BF
Applied part
Table 10-20 Isolation between Mains and Applied Limits****- Patient Leakage Main to Type CF
Applied part
Table 10-21 Total Patient leakage current: Type CF probe to ground (Only for Europe)
Table 10-22 Total Patient leakage current: Type CF probe to Mains (Only for Europe)
Conditions
Country
USA Others
Close Protective earth and close neutral with normal polarity 0.01 mA* 0.01 mA***
Close Protective earth and close neutral with reverse polarity 0.01 mA* 0.01 mA***
Open Protective earth and close neutral with normal polarity 0.05 mA* 0.05 mA***
Open Protective earth and close neutral with reverse polarity 0.05 mA* 0.05 mA***
Open neutral and close Protective earth with normal polarity 0.05 mA* 0.05 mA***
Open neutral and close Protective earth with reverse polarity 0.05 mA* 0.05 mA***
Conditions
Country
USA Others
Close Protective earth and close neutral with normal polarity 5 mA* 5 mA***
Close Protective earth and close neutral with reverse polarity 5 mA* 5 mA***
Conditions
Country
USA Others
Close Protective earth and close neutral with normal polarity 0.05 mA* 0.05 mA***
Close Protective earth and close neutral with reverse polarity 0.05 mA* 0.05 mA***
Conditions Limitation
Close Protective earth and close neutral with normal polarity 0.05 mA*****
Close Protective earth and close neutral with reverse polarity 0.05 mA*****
Open Protective earth and close neutral with normal polarity 0.1mA*****
Open Protective earth and close neutral with reverse polarity 0.1mA*****
Open neutral and close Protective earth with normal polarity 0.1mA*****
Open neutral and close Protective earth with reverse polarity 0.1mA*****
Conditions Limitation
Close Protective earth and close neutral with normal polarity 0.1 mA******
Close Protective earth and close neutral with reverse polarity 0.1 mA******