Microlab
®
STAR Service Manual
6-26
Pipetting Arm X-Alignment (continued)
Adjustment:
Before starting the Adjustment Program relieve any stress on the Pipetting Arm
by moving it several times in X-Direction
Use Service Software and let it perform approximately 10 full-length strokes in
X- direction to ensure a neutral positioning of Pipetting Arm.
Attention: since any stress on the Pipetting Arm is now relieved do not move it by
pushing or pulling it in the front. If the Pipetting Arm must be moved hold it by the
rear!