Microlab
®
STAR Service Manual
6-27
Pipetting Arm X-Alignment (continued)
• Then start "ARM X-DIFFERENCE.MCR" Adjustment Macro Program from
Service Software and carefully follow the instructions given in the program.
• Adjustment must performed if difference criteria are not met.
Interpretation of Result of Adjustment procedure:
Adjusting Pipetting Arm in X-Direction (which would be necessary in the result
window shown above).