STAR Service Manual
6-32
Corrective Actions (continued)
Attention: Do not loosen fixing screws to much, as this would lead to an uncontrolled
shift or move. A restart with the Channel Adjustment procedure is then inevitably!
-012
X- Deviation on first Level
[
X
]
Axis
X- Deviation on second Level
0022
[
Z
]
Axis
If no Channel positioning
Tool P/N 182960 is
available then use upper
Fixing screw as pivot point
and tilt Channel
counterclockwise.