Microlab
®
STAR Service Manual
6-33
Corrective Actions (continued)
Using the Channel positioning Tool P/N 182960:
Note: Y-Spindle may remain in place.
Place the Channel positioning Tool onto
Pipetting Arm and move it in Y Direction
towards the channel to be corrected.
The goal is to take over the current
position of the channel in question and
freeze its position with the Channel
positioning Tool.
After the Channel positioning Tool is
attached to the channel, start tightening it.
The clamps of the Channel positioning
Tool should hold the channel tight, then fix
the Channel positioning Tool with the two
knurled fixing screws onto the rail.
Clamps -
and
Knurled fixing screws of positioning Tool.