Heart and Respiration Rate Monitoring System powered by EarlySense
98 Centrella® Smart+ Bed with Heart and Respiration Rate Monitoring System powered by EarlySense Instructions for Use
(193587 REV 10)
The HR or RR alarm can be silenced or suspended. The Silence or
Suspend control allows the caregiver to temporarily stop alarms from
being sent to the bed and/or nurse call system, if connected.
When the Suspend control is pressed, these will occur:
• The yellow alarm message will show and suspends the alarms
for 15 minutes.
• A timer will show at the bottom of the touchscreen that will
countdown from the time of the alarm.
After 15 minutes, Suspend mode will stop and the system will resume monitoring the HR and RR.
High or Low RR Alarm—a message shows when the respiration rate is
above or below the high or low alarm threshold settings.
High or Low HR Alarm—a message shows when the heart rate is
above or below the high or low alarm threshold settings.
Lost Patient Signal—this message shows when the sensor can not
detect the patient signals. Check the patient and sensor location.
Unstable Signal—this message shows when the sensor can not
continually detect the patient. Consider using an alternative
monitoring device.
Bed Sensor Error—this message shows when the sensor
malfunctioned or the sensor is disconnected. Make sure the sensor
cable is fully connected to its port on the back of the head deck panel.
If you replaced the sensor, make sure the activation code was entered
correctly.