Graphical Caregiver Interface (GCI)® Controls
54 Centrella® Smart+ Bed Instructions for Use (193587 REV 10)
Re-Zero (Same Patient)
NOTES:
• If the “Not In Required Position” screen shows, press the
Adjust Bed Button on the touchscreen.
• If the “Difference from Last Zero” caution screen shows, re-
check the bed for items that should be included during
the zero.
• The scale will not zero if there is more than a 45 kg (99.2 lb)
weight difference from the last zero. This is to reduce the
possibility of zeroing the scale with a patient in the bed.
4. A single beep will sound and the touchscreen will show that the
bed is ready for a new patient. To return to the Home screen, touch
the screen or place a patient in the bed.
To learn more about the Foley Position Limit, see “Foley Position Limit”
on page 72.
Use Re-Zero whenever you add equipment or other items to a bed that previously has been zeroed for
the patient. An example of this would be if you replace a mattress with a heavier mattress or if there are
inconsistent weight readings. Before you press Re-Zero, make sure that the patient is not in the bed.
The Re-Zero control will—
• Not reset the bed settings.
• Zero the Scale for the Same Patient (does not clear the history).
1. Press Re-Zero.
2. A reminder message shows. Follow the on-screen instructions, and
press Continue.
Cancel will return the touchscreen to the previous screen.
3. The touchscreen will show that the bed is zeroing. Do not touch the
bed while the bed zeroes.
NOTES:
• If the “Not Required Position” screen shows, press the Adjust Bed Button on the
touchscreen.
• If the “Difference from Last Zero” caution screen shows, re-check the bed for items that
should be included during the zero.
• The scale will not zero if there is more than a 26 kg (57.3 lb) weight difference from the last
zero. This is to reduce the possibility of zeroing the scale with a patient in the bed.