Heart and Respiration Rate Monitoring System powered by EarlySense
84 Centrella® Smart+ Bed with Heart and Respiration Rate Monitoring System powered by EarlySense Instructions for Use
(193587 REV 10)
GCI Symbols—HR/RR Monitoring System powered by EarlySense
HR/RR MONITORING SYSTEM POWERED BY EARLYSENSE SPECIFICATION
HR AND RR MONITORING SYSTEM POWERED BY EARLYSENSE SAFETY INFORMATION
Read all precautionary information and specifications before use.
Symbol Description Page Symbol Description Page
Heart Rate—ON 94 Out of Room Alarm—ON indi-
cator
100
Respiration Rate—ON 94 Out of Room Alarm—OFF
indicator
100
Heart Rate—OFF 99 In-room Alarm—OFF indica-
tor
99
Respiration Rate—OFF 99 In-room Alarm—ON indicator 99
Unstable signal 95 Lost patient signal 95
Service required 95 Multiple Alarms (HR, RR, Ser-
vice, or HR/RR monitoring
sensor)
95
Resume Monitoring 97 Suspended Alarm 97
Heart Rate Respiration Rate
Range 30 - 170 beats per minute (BPM) 6 - 45 breaths per minute (Br/min)
Average Period 1 minute 1 minute
Accuracy +/- 4% or +/- 5 BPM whichever is
greater
+/- 4% or +/- 1.5 Br/min whichever is
greater
The Heart and Respiration Rate Monitoring System powered by EarlySense is contraindicated for use
in—
• Contraindication—Patients whose proper positioning can not be achieved or maintained.
• Contraindication—Patients who do not meet the weight limits tested or specified.
• Contraindication—Situations where a dry environment can not be ensured.
• Contraindication—An MRI environment.