Heart and Respiration Rate Monitoring System powered by EarlySense
Centrella® Smart+ Bed with Heart and Respiration Rate Monitoring System powered by EarlySense Instructions for Use 85
(193587 REV 10)
• Contraindication—An explosive atmosphere or in the presence of flammable anesthetics or
gases.
To prevent injury or equipment damage obey these warnings:
• Warning—Read and understand all warnings on the unit itself prior to use with a patient.
• Warning—The Centrella® Smart+ Bed with or without the Heart and Respiration Rate
Monitoring System powered by EarlySense, has an intended patient range of 70 lb (32 kg) to
500 lb (227 kg). Use of the bed with patients outside the intended patient weight range may
increase the risk of entrapment, asphyxiation, and/or loss of therapeutic value. Use of the bed
with patients above the bed’s intended maximum patient weight may also cause damage the
bed.
• Warning—Follow the product manufacturer’s instructions.
• Warning—The potential for electrical shock exists with electric equipment. Failure to follow
facility protocols may cause death or serious injury.
• Warning—To avoid risk of electric shock, this equipment must only be connected to a supply
mains with protective earth.
• Warning—Make sure to use a correctly grounded, three-prong, 120 V outlet (NEMA 5-15R or
NEMA 5-200R outlet, rated 125 V AC, 15 A or 125 V AC, 20 A, respectively). Failure to do so
could cause personal injury, fire, or damage to the equipment or facility wiring.
• Warning—Connect the power cord to hospital grade receptacles only.
• Warning—Incorrect use or handling of the power cord may cause damage to the power cord.
• Warning—Take care to minimize the risk of tripping over the power cord by carefully locating
it from the unit to its power source.
• Warning—Make sure the position of the unit is such that you can quickly, without
obstruction, unplug the power cord from the main power supply if necessary.
• Warning—Do not use the bed in an oxygen rich environment or with oxygen tents.
• Warning—Do not operate the bed in the presence of flammable gas or vapors.
• Warning—Do not operate the bed in the presence of flammable anesthetics or nitrous oxide.
• Warning—Before maintenance or service is done on the unit, make sure to unplug the unit
and remove the patient or fully remove the unit from use.
• Warning—Operate the bed within the stated environmental conditions; see “Environmental
Conditions for Use” on page 134.
• Warning—This device is not compatible for use in Magnetic Resonance Imaging (MRI).
• Warning—Do not connect the power cord to an extension cord or multiple outlet strip. There
is a risk of overheating and fire may cause injury or damage.
• Warning—Connect only items that have been specified as parts of the device or compatible
with the device.
• Warning—Consult your local regulations to safely discard or recycle electronic equipment
and batteries.
• Warning—Do not discard as unsorted municipal waste. See your local distributor for
collection and/or recycling system available in your country.
• Warning—Federal USA law restricts this device to sale by or on the order of a physician.
• Warning—The data acquired by HR and RR Monitoring System powered by EarlySense should
be interpreted by a Healthcare Practitioner only.