Transport The Patient
Centrella® Smart+ Bed Instructions for Use Manual (193587 REV 10) 111
WARNING:
Warning—Failure to remove power from the bed could cause injury or equipment damage.
5. Unplug the bed.
6. Lower the siderails.
7. Disconnect the mattress’s four attachment knobs from the bed frame.
8. If the bed has the pro+ mattress (P7923A), see the pro+ Mattress Instructions for Use (209196) to
remove the mattress. If the bed has the Centrella® max mattress (P7922A), do as follows. Otherwise,
go to
Step 9.
9. Remove the mattress from the bed.
TRANSPORT THE PATIENT
TRANSPORT SAFETY INFORMATION
WARNING:
To help prevent injury and/or equipment damage, obey these warnings when you transport a patient
in the bed:
• Warning—If the bed is stopped on a ramp, set the brake to avoid unwanted bed movement.
• Warning—During transport, use caution so that the bed does not tip or overbalance.
a. Fold the foot end of the mattress assembly over the head end.
b. Pull up on the heel cover (A) to release it,
and then turn it as necessary to fully
disconnect the cover from its opening.
c. Pull the cover (A) back toward the sleeve to
get access to the pneumatic box
connection.
d. Disconnect the interface connector
assembly (B) from the pneumatic box.
NOTE:
To disconnect the interface connector assembly,
it may be helpful to insert a small screwdriver
between the connector assembly and its latch
tabs.
CAUTION:
Caution—Failure to keep the pneumatic box
covered when the interface connector assembly is
disconnected could cause equipment damage.
e. Pull the pneumatic box cover (C) over the
pneumatic box (D). Keep the interface
connector on the pneumatic box covered
until you connect the interface connector
assembly.