Alarm, Alerts, and Information Indicators
Centrella® Smart+ Bed Instructions for Use (193587 REV 10) 13
LOW PRIORITY ALARM
If the HR and RR Monitoring System option is active, a two tone alarm (e - c) will sound every 3 seconds
to indicate that a HR or RR alarm is active and a priority nurse call is sent to the nurses station (for beds
equipped and connected to a Nurse Call System).
BEEP ALERTS
A single beep will sound when an activity is successful.
A triple beep will sound to indicate that you should look at the touchscreen, attention is needed.
For a critical fault, a triple beep will sound every 10 seconds until the user acknowledges the error on the
touchscreen. Follow the on-screen instructions. If the critical fault is not resolved, the triple beep will
sound again after an hour. The cycle will continue until the issue is resolved.
A low volume repeating triple beep every 10 seconds will sound to remind you that the Bed Exit is
silenced and should be resumed before you leave the room.
A quadruple beep (four beeps) will sound to indicate a Bed Exit alert.
A continuous tone will sound to indicate a serious bed malfunction and that immediate attention is
needed.
VERBAL ALERTS
These Verbal Alerts are available on some models:
• “Please don’t get up.”
• “The Care Team has been called.”
•“Brake not set.”
•“Function unavailable.”
• “Call light not connected.”
• “Obstacle detected!”
The Verbal Alerts can be disabled through the touchscreen except for the “Brake Not Set” alert (see
“Voice Alerts” on page 70). The “Brake Not Set” verbal alert can not be disabled.
BED EXIT INDICATOR
ALERT SILENCE INDICATOR
The Bed Exit indicator is a visual indicator on the caregiver control panel that will be
green when the Bed Exit feature is on.
The Bed Exit Alert Silence indicator is a visual indicator on the caregiver control panel.
When the Alert Silence indicator is on, the bed stops monitoring patient movement
temporarily (30 seconds default).
This will only silence Bed Exit Alerts.