EasyManua.ls Logo

Hillrom Centrella Smart+ Bed - (Change Zero Items)

Hillrom Centrella Smart+ Bed
160 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Graphical Caregiver Interface (GCI)® Controls
Centrella® Smart+ Bed Instructions for Use (193587 REV 10) 61
+/- (Change Zero Items)
3. Press Re-Zero.
4. A reminder message shows. Follow the on-screen instructions, and
press Continue.
NOTE:
Cancel will return the touchscreen to the previous screen.
5. The touchscreen will show that the bed is zeroing/taring. Do not
touch the bed while the bed zeroes/tares.
NOTES:
If the “Not Required Position” screen shows, press the Hold to Adjust control on the
touchscreen.
If the “Difference from Last Zero” caution screen shows, re-check the bed for items that
should be included during the zero.
The scale will not zero if there is more than a 26 kg weight difference from the last zero.
This is to reduce the possibility of zeroing the scale with a patient in the bed.
6. A single beep will sound and the screen will show that the Re-Zero
is complete.
The +/- control allows you to manually change the weight for items
added to or removed from the bed to correct the patient weight while
the patient is on the bed.
NOTE:
If the patient is not on the bed, use the Re-Zero control.
Before you add or remove items, use the +/- control to keep the patient weight. Then, add an item to or
remove an item from the bed so the item is not included in the patient weight measurement.