Loading...
Alarm, Alerts, and Information Indicators
Centrella® Smart+ Bed Instructions for Use (193587 REV 10) 15
Powered Transport Battery Charge Indicator
NOTE:
The battery does not support these: Patient controls, Auxiliary Outlet, Scale, Bed Exit Alert System, HR
and RR monitoring, or the air support system.
The battery stays engaged for 1 minute after the last control is pressed, then goes into a sleep mode.
If the Battery Indicator changes from Charged to Low consistently within four hours of being
unplugged from AC power, replace the battery.
If the battery has no charge, it may take up to 24 hours to charge to operational status.
To make sure the battery is always charged, plug the bed into an AC power outlet whenever possible.
There are two bed frame battery charge indicators on this bed:
Caregiver Control Panel—Bed Battery
Charge Indicator
Touchscreen—Bed Battery Charge
Level Indicator
Caregiver Control Panel—Bed Frame Battery Charge Indicator
The battery backup indicator on the caregiver control panel shows the battery status:
On = The battery is charging.
Flashing =The bed is powered by battery backup, or the battery needs to be charged. A
triple beep sounds every two minutes when the battery reaches a low condition and the bed
is unplugged.
Off = The battery is not engaged, or the battery charge is critically low and can not operate
the motors.
Touchscreen Bed Frame Battery Charge Level Indicator
The battery charge status shows at the top of the touchscreen.
The screen will dim 10 seconds after the last control is pressed,
then will go to sleep after 30 seconds. The screen will come on
when AC power is restored to the bed
Low—the Low indicator (one yellow bar) flashes when the battery charge is low. A triple beep
sounds every two minutes when the battery reaches a low condition and the bed is unplugged.
CAUTION:
Caution—For transport, a fully-charged battery is preferred. If the battery charge is low, before you
unplug the bed, put the bed in the transport position before any transport. Connect the bed to AC