Centrella® Smart+ Bed Instructions for Use Manual (193587 REV 10) 135
AC/Mains Power Requirements
Auxiliary Outlet Power Specifications
There are no user accessible fuses. Refer to the Centrella® Smart+ Bed Service Manual (193588) for fuse
ratings and replacement procedures.
Applied Parts (in accordance with IEC 60601-1)
Bed Classification and Standards
Flammability Codes—United States and Canada
All recommended mattresses meet the applicable United States and Canadian flammability
specifications (see the product’s law label).
Condition Range
Rated voltage 100-240 V AC
Input current rating 10 A
Frequency 50-60 Hz
Fuse Type
Receptacle 10 A outlet, electrically isolated from the bed’s
mains power (120 V)
Applied parts Applied parts
Siderail Headboard
Footboard Patient pendant
Sleep deck Mattress
Classification Standard
Technical and quality assurance standards AAMI ES 60601-1
IEC 60601-1
IEC 60601-2-52
IEC 60601-2-49
IEC 60601-1-8 (for beds with HR and RR Monitor-
ing System
a
)
CAN/CSA-C22.2 No. 60601-1
ISO13485
a. The IEC 60601-1-8 standard only applies to the Heart and Respiration Rate Monitoring System powered by
EarlySense.
Equipment classification per EN 60601-1 Class 1
Degree of protection against electric shock Type B
Degree of protection against ingress of water IPX4
Degree of protection against the presence of
flammable anesthetic mixtures
Not for use with flammable anesthetics
Mode of operation Non-continuous: 2 min on / 18 min off duty cycle
Sound level <65 dBA
Application environments Application Environment 1,2,3, and 5 per EN
60601-2-52