6.3 Hold Function
______________________________________________________________________________________________
Distinction between the PASS Hold Function, FAIL Hold Function,
and Hold Function
With a comparative-voltage value set in withstand voltage test, if the
output voltage deviates from the comparative-voltage range, the unit
switches to the FAIL state.
If the test time is set to OFF in withstand voltage mode, PASS screening
is not performed. In such a case, FAIL screening is performed or the test
is terminated using the STOP key.
In insulation resistance mode, if the test time is set to OFF, PASS
screening is not performed. In such a case, the test is terminated using the
STOP key.