10.2 General Specifications
______________________________________________________________________________________________
Display Fluorescent tube display (digital display)
Monitor function Output voltage / Current / Measurement resistance
Monitor cycle 2 Hz or faster
Ambient operating
conditions
0to40 (32 to 104 ), 80%RH or less (no condensation)
Ambient storage
conditions
-10 to 50 (14 to 122 ), 90%RH or less (no condensation)
Ambient conditions for
assured accuracy
Accuracy assurance
period
23 5 (73 9 ), 80%RH or less (no condensation)
after 5 minutes minimum warm-up
1 year
Operating Envirowment Indoors. <2000 mASL (6562 feet)
Power supply 100 VAC (3159)
120 VAC (3159-01)
220 VAC (3159-03)
230 VAC (3159-04)
240 VAC (3159-05)
(Voltage fluctuations of 10% from the rated supply voltage are taken into
account.)
Rated power frequency 50/60 Hz
Dielectric strength 1.35 kVAC, 10 mA, 1 minute between power supply and frame
Maximum rated power 800 VA
Dimensions Approx. 320W x 155H x 330D mm (12.60"W x 6.10"H x 12.99"D)
(excluding projections)
Mass Approx. 18 kg (634.9 oz.) (3159)
Approx. 20.5 kg (88.2 oz.) (3159-01)
Approx. 21.5 kg (758.4 oz.) (3159-02/-03/-04)
Fuse 250VT8AL (3159, 3159-01), 250VT4AL (3158-02/03/-04)
Accessories 9615 H.V. TEST LEAD (high voltage side and return, 1 each)
Power cord
Instruction manual
Spare fuse
Options 9613 REMOTE CONTROL BOX (SINGLE)
9614 REMOTE CONTROL BOX (DUAL)
9267 SAFETY TEST DATA MANAGEMENT SOFTWARE
9637 RS-232C CABLE (9pin-9pin/ 1.8 m)
*
9638 RS-232C CABLE (9pin-25pin/ 1.8 m)
*
* Doesn't comply with CE mark requirements when used with the 3159.
Standard Applying
EMC
Safety
EN61326-1:1997+A1:1998 CLASS A
EN61000-3-2:1995+A1:1998+A2:1998
EN61000-3-3:1995
EN61010-1:1993+A2:1995
Power supply: Pollution Degree 2, overvoltage category II
(anticipated transient overvoltage 2500 V)
EN61010-2-031:1994