6.12 Example of Optional Functions Use
______________________________________________________________________________________________
Optional function settings
Optional Functions Selection
PASS Hold Function
1
: Held
FAIL Hold Function
1
: Held
Optional function settings
Optional Functions Selection
FAIL Hold Function
1
: Held
Momentary Out
1
:Set
FAIL mode
1
:Set
6.12 Example of Optional Functions Use
The following describes how 3159 optional functions are used for testing.
Various combinations of optional functions are possible for testing.
(1) Testing to check test results
Advantages of these settings
PASS or FAIL state is held, allowing inspection of test results.
(2) Safe testing by remote control
Advantages of these settings
Hold down the START key during the test, as the Momentary-Out function
is set. The 9614 REMOTE CONTROL BOX (DUAL) must be operated
with both hands during the test. This prevents high-voltage devices such
as the probe and tested device from coming into contact with the hands.
The FAIL Hold function must be cancelled using the STOP keyonthe
unit, as the FAIL mode is set. The use of the FAIL mode enables the FAIL
state to be set.