27
_____________________________________________________________________________________________
3.3 "SETTING " State
______________________________________________________________________________________________
3.3.3 Setting the Test Time
Test Time:
60.0 s
Test Time:
OFF
NOTE
The setting resolution of the test time is 0.1 s between 0.5 and 99.9 s, and
becomes1sbetween100and999s.
If a test time has been set, the reduction timer will operate during the test.
If the test time is set to OFF, the comparative-voltage value becomes ineffective.
If the test time is set to OFF, it facilitates output-voltage settings.
If the test time is set to OFF, the time elapsed during the test is displayed. When
this time exceeds 999 s
,
"---" will a
pp
ear
,
but the test will continue.
1.
If no flashing cursor is displayed in the withstand voltage mode READY
state, press either the / key to display the cursor in the test time
position.
2. Set the test time using the
/ keys.
With time set the time changes in 0.1 s increments (1 s increments when
the set time scale is 100 to 999 s).
With time set at 0.5 s to 99.9 s, press SHIFT + / keys. The time
changes in 1.0 s increments (10 s increments when the set time scale is
100 s to 999 s).
Settings may be made along a scale ranging from 0.5 s to 999 s.
If no testing time is required, turn off the ON/OFF key.
3. To finalize the test settings, press the
STOP key. The settings are
confirmed and the unit returns to the READY state.