130
_____________________________________________________________________________________________
9.4 Command
______________________________________________________________________________________________
9.4.1 Common Command Messages
Syntax
∗
IDN?
Response
syntax
<data>
<data>
Manufacturer's name, Model name, Serial number (Not used - always zero), Software version
Function
Queries manufacturer's name, model name, and software version.
Example
Queries manufacturer's name, model name, and software version.
Transmission
∗
IDN?
Response HIOKI,3159,0,V01.00
Syntax
∗
RST
Response
syntax
OK
Resetting completion
Function
Resets the 3159. The items which are reset are listed below.
Withstand voltage mode (initial setting)
Comparative voltage value setting: OFF
Test time setting: ON
Lower limit value setting: OFF
Comparative voltage value: 0.00 kV
Upper limit value: 0.2 mA
Lower limit value: 0.1 mA
Test time: 0.5 s
Output voltage range: 2.5 kV
Insulation resistance mode (initial setting)
Test time setting: ON
Upper limit value setting: OFF
Test voltage value: 500 V
Upper limit value: 2000 mΩ
Lower limit value: 1000 mΩ
Test time: 0.5 s
Optional Functions
0.1.0.0 0.0.0.0 0.0.0 (ON: FAIL hold only)
Example
Performs device initial setting.
Transmission
∗
RST
Performs device initial setting.
Response OK
Resetting completion
∗IDN?
Queries manufacturer's name, model name, and software version.
∗RST
Performs device initial setting.