11.3 Troubleshooting
______________________________________________________________________________________________
Symptom
What to check and Solution
The screen does not
illuminate when the
power is turned on.
Is the power cord disconnected?
Connect the power cord.
Has the fuse blown?
Replace the fuse.
The test will not begin
even if the
START
key is pressed.
Is
READY
lit?
Is
EXT
lit?
The external I/O and the REMOTE CONTROL BOX
have priority over the unit's
START
key. Turn off the
power to disable the external I/O and the REMOTE
CONTROL BOX, then turn on the power.
Is Double Action set in Optional Functions?
IfDoubleActionisset,pressthe
STOP
key first and
then press the
START
key.
11.3 Troubleshooting
If damage is suspected, check the "Troubleshooting" section before
contacting your dealer or Hioki representative.
If any of the following should occur, stop using the unit, disconnect the
power cord and probe, and contact your dealer or HIOKI representative.
If you are certain that the unit is damaged.
If the measurement you wish to perform is inoperative.
If the unit was stored for a long period of time in high temperatures and
humidity, or other undesirable conditions.
If the unit was damaged in transit.